Teseq NSG 4070

From RadiWiki
Jump to: navigation, search

The Teseq NSG 4070 device driver is a Signal GeneratorAmplifierCouplerPower MeterAD Converter which is supported by RadiMation®.

Links

This information about an equipment is automatically generated from the list of supported device drivers from RadiMation.
If you need more information you can contact radimation-support@raditeq.com.

How to install and connect a NSG 4070 to Radimation over USB.

These amplifiers have diverse protocols to connect.

These are LAN (TCP/IP), RS232 and USB

This information regards the connection via USB. If you connect the device via USB it is important to know that it is RS232 over USB. In the documentation of the amplifier it is not very clear, however it is needed to understand and helps to configure the device in Radimation.

  • Make sure that the device is set to USB. You can find this via the setup button on the device itself. Please reerence te manual from Teseq for further information.
  • Connect the device physically via USB.
  • Open device manager (Type device manager in the search window next to start in windows)

Device manager.png

  • Open your com ports and look for the name NSG 4070 (The picture above is a example. When the NSG4070 is present and recognised a com port with the name NSG 4070 should be visible here. If not please reference the manual from the device)
  • Check which com port this is
  • Open Radimation
  • Go to configuration/configuration/device drivers/amplifiers

NSG4070.png

  • Add the driver for the NSG4070 by clicking add
  • Open the settings by double clicking
  • Click "advanced"
  • Go to the tab "communication"
  • Set "Communication streams" to: RS232
  • Set the COM port to the number that you have found in device manager from Windows
  • Set the Baud rate to 9600 baud !! (this is important, it is not in the documentation of the manufacturer. If you want to use the USB connection then it's communication is using RS-232 with a fixed baud rate of 96000)
  • Leave the rest default.
  • Click "Ok" and press "Check"
  • If all was set correctly you will receive a message that the communication was successful. The amplifier is now configured and ready for use in the software