RadiMation Error: 805328368

From RadiWiki
Revision as of 12:27, 25 March 2020 by Tjsc (talk | contribs) (Ch. 1: Binary block buffer couldn't be found)
Jump to: navigation, search

Ch. 1: Binary block buffer couldn't be found

This a known issue when using the LecroyWave Surfer en waverider. Make sure that the device is in LXI mode when using it with RadiMation. When you get the Error "Binary Block buffer can't be found" it could indicate that the device is set to TCP/IP mode. Please verify if this is the case. Set the communication protocol to LXI and then try to connect again. When the device is configured via TCP/IP it will respond correctly to the check devices test but it will give an error when running the test.

For further information please read this form post regarding the use of TCP/IP, LXI and the Lecroy waverider in combination with different software. Post Usage LXI/TCP-IP

For more information on how to change the communication protocol on the device please check the manual of the device.

When you have changed the protocol to LXI and are still having issue please contact our support for further assistance.