Difference between revisions of "RadiMation Error: 805328368"

From RadiWiki
Jump to: navigation, search
(Created page with "== Ch. 1: Binary block buffer couldn't be found == This a known issue when using the LecroyWave Surfer en waverider. Make sure that the device is in LXI mode when using it wi...")
 
(Ch. 1: Binary block buffer couldn't be found)
Line 5: Line 5:
  
 
For further information please read this form post regarding the use of TCP/IP, LXI and the Lecroy waverider in combination with different software.
 
For further information please read this form post regarding the use of TCP/IP, LXI and the Lecroy waverider in combination with different software.
[
+
[https://forums.ni.com/t5/LabVIEW/Lecroy-Waverunner-TCP-IP-vs-LXI/td-p/1584422 Post Usage LXI/TCP-IP]
https://forums.ni.com/t5/LabVIEW/Lecroy-Waverunner-TCP-IP-vs-LXI/td-p/1584422 Post Usage LXI/TCP-IP in combination with a Lecroy WaveRunner on the NI forum, the information is also valid for the wave surfer]
+
 
 +
For more information on how to change the communication protocol on the device please check the manual of the device.
 
   
 
   
 
When you have changed the protocol to LXI and are still having issue please contact our support for further assistance.
 
When you have changed the protocol to LXI and are still having issue please contact our support for further assistance.

Revision as of 12:27, 25 March 2020

Ch. 1: Binary block buffer couldn't be found

This a known issue when using the LecroyWave Surfer en waverider. Make sure that the device is in LXI mode when using it with RadiMation. When you get the Error "Binary Block buffer can't be found" it could indicate that the device is set to TCP/IP mode. Please verify if this is the case. Set the communication protocol to LXI and then try to connect again. When the device is configured via TCP/IP it will respond correctly to the check devices test but it will give an error when running the test.

For further information please read this form post regarding the use of TCP/IP, LXI and the Lecroy waverider in combination with different software. Post Usage LXI/TCP-IP

For more information on how to change the communication protocol on the device please check the manual of the device.

When you have changed the protocol to LXI and are still having issue please contact our support for further assistance.